11.1.5 Memory
If Memory is selected then the transmitter will perform a RAM and ROM memory test. Test pat-
terns will be written to and read from all RAM memory locations. The ROM checksum will be
recalculated and compared to the value stored in the ROM.
11.1.6 Set Relay
The Set Relays diagnostic menu allows for the activation/deactivation of each
Relay. To access relays 5 and 6 on dual channel transmitters, press [ENTER].
0 = Normal (normally open contacts are open)
1 = Inverted (normally open contacts are closed)
Press [ENTER] to return to Measurement mode.
11.1.7 Read Relays
The Read Relays diagnostic menu shows the state of each Relay as defined below. Press
[ENTER] to exit from this display.
0 = Normal
1 = Inverted.