Instruction Manual
pH 1100/ pH 2100
Bench pH/Ion Meter
Technolo
g
y
M
ade
E
a
s
y
.
.
.
68X090818
rev 2.2 Nov 02
Page view 0
1 2 3 4 5 6 ... 63 64

Summary of Contents

Page 1 - Instruction Manual

Instruction Manual pH 1100/ pH 2100 Bench pH/Ion Meter TechnologyMadeEasy...68X090818 rev 2.2 Nov 02

Page 2

Instruction Manual CyberScan pH 1100/ 2100 10 2 STARTING UP Attention: Do not get water on the BNC connector during operation. Avoid touching th

Page 3 - PREFACE

Instruction Manual CyberScan pH 1100/ 2100 11 3 PH CALIBRATION & MEASUREMENT 3.1 pH calibration This meter is capable of up to 5 point calibra

Page 4 - TABLE OF CONTENTS

Instruction Manual CyberScan pH 1100/ 2100 12 3.1.2 Starting pH calibration 1. Turn the meter ON. If the meter is not in the pH measurement mode

Page 5

Instruction Manual CyberScan pH 1100/ 2100 13 3.1.3 Standard pH buffer calibration 1. When entering the standard calibration mode, the first buf

Page 6 - 1 INTRODUCTION

Instruction Manual CyberScan pH 1100/ 2100 14 3.1.4 Custom pH buffer calibration 1. When entering the custom calibration mode, the lower display

Page 7 - Figure 1: Membrane Keypad

Instruction Manual CyberScan pH 1100/ 2100 15 3.1.5 Calibration error messages Standard pH Buffer Calibration: If the first buffer value of 7.0

Page 8 - 1.3 Rear Panel

Instruction Manual CyberScan pH 1100/ 2100 16 3.2 pH measurement This meter is capable of taking measurements with automatic or manual temperature

Page 9 - 1.4 Electrode Holder

Instruction Manual CyberScan pH 1100/ 2100 17 3.2.3 Taking pH Measurements NOTE: Be sure to remove the electrode soaker bottle or protective rub

Page 10 - Attention:

Instruction Manual CyberScan pH 1100/ 2100 18 4 mV CALIBRATION AND MEASUREMENT 4.1 mV Calibration 1. Turn the meter ON. If the meter is not i

Page 11 - Selectable buffer sets:

Instruction Manual CyberScan pH 1100/ 2100 19 4.2 mV Measurement 1. Turn the meter ON. If the meter is NOT in the mV measurement mode, press EXIT

Page 13 - Cal Done

Instruction Manual CyberScan pH 1100/ 2100 20 5 ION CALIBRATION AND MEASUREMENT (ONLY APPLICABLE FOR pH 2100) The pH 2100 meter reads ion concent

Page 14 - NOTE:

Instruction Manual CyberScan pH 1100/ 2100 21 If you are performing a 2-point calibration, press EXIT. The screen flashes Aver. Slope = (the aver

Page 15 - . See Figure 26. Press

Instruction Manual CyberScan pH 1100/ 2100 22 5.3 Ion measurement The pH 2100 meter features two different modes of ion measurement: high resoluti

Page 16 - 3.2 pH measurement

Instruction Manual CyberScan pH 1100/ 2100 23 6 MEMORY FUNCTIONS 6.1 Memory input This meter can store up to 100 sets of data in several combinat

Page 17 - Figure 36

Instruction Manual CyberScan pH 1100/ 2100 24 6.2 Memory recall This function recalls the previous readings stored in the memory. You can only acc

Page 18 -

Instruction Manual CyberScan pH 1100/ 2100 25 7 STABILITY INDICATOR The stability indicator signals when pH, mV or Ion readings are stable. The pH

Page 19 - pH/Temp mV Ion

Instruction Manual CyberScan pH 1100/ 2100 26 9 TEMPERATURE CALIBRATION 9.1 Temperature Calibration For this calibration procedure, the ATC probe

Page 20 - 5.1 Ion calibration

Instruction Manual CyberScan pH 1100/ 2100 27 10 SETUP MODE The advanced setup mode lets you customize your meter’s preferences and defaults. The

Page 21 - Sys Updating. .

Instruction Manual CyberScan pH 1100/ 2100 28 10.1.2 mV setup submenus Press SETUP to enter setup mode. Use >/CAL to select mV, and then pres

Page 22 - 5.3 Ion measurement

Instruction Manual CyberScan pH 1100/ 2100 29 10.1.4 Meter general configuration setup submenus Press SETUP to enter setup mode. Use >/CAL to

Page 23 - Data Log. . . . Loc: 9

CyberScan pH 1100/ 2100PREFACE Thank you for selecting the pH 1100/ 2100 benchtop meter. This meter measure pH, millivolts, relative millivolts and t

Page 24 - 6.2 Memory recall

Instruction Manual CyberScan pH 1100/ 2100 30 10.2 pH/ temperature setup mode 10.2.1 Entering pH/temperature setup mode 1. Turn the meter ON.

Page 25 - 7.546pH 25.4 C ATC

Instruction Manual CyberScan pH 1100/ 2100 31 10.2.3 pH resolution setup program P1.1 Use this program to select pH resolution (0.01 or 0.001 pH)

Page 26 - 9.1 Temperature Calibration

Instruction Manual CyberScan pH 1100/ 2100 32 10.2.5 pH Measurement Alarm Setup Program P1.3 Use this program to select alarm for low and high pH

Page 27 - 10.1 Setup mode overview

Instruction Manual CyberScan pH 1100/ 2100 33 10.2.6 pH Calibration Due Alarm Setup program P1.4 Use this program to set a reminder when the next

Page 28

Instruction Manual CyberScan pH 1100/ 2100 34 10.2.7 Stability Setup Program P1.5 Use this program to turn on or off the pH stability indicator a

Page 29 - Press ENT. to Select

Instruction Manual CyberScan pH 1100/ 2100 35 10.2.8 View pH calibration data setup program P1.6 Use this program to view the following: • Last

Page 30 - Select & Press ENTER

Instruction Manual CyberScan pH 1100/ 2100 36 10.2.9 pH calibration data reset setup program P1.7 Use this program to erase all pH calibration da

Page 31 - Figure 86

Instruction Manual CyberScan pH 1100/ 2100 37 10.3 mV setup mode 10.3.1 Entering mV setup mode 1. Turn the meter ON. If the meter is not in th

Page 32 - Key In High Alarm

Instruction Manual CyberScan pH 1100/ 2100 38 10.3.2 mV measurement alarm setup program P2.0 Use this program to select alarm for low and high mV

Page 33 - Figure 93

Instruction Manual CyberScan pH 1100/ 2100 39 10.3.3 mV calibration due alarm setup program P2.1 Use this program to set a reminder when the next

Page 34 - Figure 96

Instruction Manual CyberScan pH 1100/ 2100 4 TABLE OF CONTENTS 1 INTRODUCTION 6 1.1 Introducing the Bench Meter Series 6 1.2 Keypad 7 1.3 Rear

Page 35 - = 0.2 mV

Instruction Manual CyberScan pH 1100/ 2100 40 10.3.4 Stability setup program P2.2 Use this program to turn on or off the mV stability indicator a

Page 36 - Figure 108

Instruction Manual CyberScan pH 1100/ 2100 41 10.3.5 View mV calibration data setup program P2.3 Use this program to view the following: • Last m

Page 37 - Figure 111

Instruction Manual CyberScan pH 1100/ 2100 42 10.3.6 mV calibration data reset setup program P2.4 Use this program to erase all mV calibration da

Page 38 - Figure 115

Instruction Manual CyberScan pH 1100/ 2100 43 10.4 Ion setup mode (pH 2100 meter only) 10.4.1 Entering ion setup mode 1. Turn the meter ON. If

Page 39 - Figure 118

Instruction Manual CyberScan pH 1100/ 2100 44 10.4.3 Ion measurement alarm setup program P3.1 Use this program to select alarm for low and high i

Page 40 - Figure 121

Instruction Manual CyberScan pH 1100/ 2100 45 10.4.4 Ion calibration due alarm setup program P3.2 Use this program to set a reminder when the nex

Page 41 - = 3.2 mV

Instruction Manual CyberScan pH 1100/ 2100 46 10.4.5 Stability setup program P3.3 Use this program to turn on or off the ion stability indicator

Page 42 - Figure 128

Instruction Manual CyberScan pH 1100/ 2100 47 10.4.6 Ion mode setup program P3.4 Use this program to select high or low ion resolution mode. High

Page 43 - Figure 133

Instruction Manual CyberScan pH 1100/ 2100 48 10.4.7 View ion calibration data setup program P3.5 Use this program to view the following: • Last

Page 44 - Figure 137

Instruction Manual CyberScan pH 1100/ 2100 49 10.4.8 Ion calibration data reset setup program P3.6 Use this program to erase all ion calibration

Page 45 - Figure 140

Instruction Manual CyberScan pH 1100/ 2100 5 10.2.9 pH calibration data reset setup program P1.7 36 10.3 mV setup mode 37 10.3.1 Entering mV setu

Page 46 - Figure 143

Instruction Manual CyberScan pH 1100/ 2100 50 10.5 Meter general configuration setup mode 10.5.1 Entering meter general setup mode 1. Turn the

Page 47 - Ion Mode Setup : P3.4

Instruction Manual CyberScan pH 1100/ 2100 51 10.5.3 Backlight setup mode P4.1 Use this program to turn backlight on or off and set duration time

Page 48

Instruction Manual CyberScan pH 1100/ 2100 52 10.5.4 Data log setup mode P4.2 Use this program select automatic or manual data logging. Automatic

Page 49 - Figure 158

Instruction Manual CyberScan pH 1100/ 2100 53 10.5.6 Communication setup mode P4.4 Use this program to set parameters of serial communication. 1

Page 50

Instruction Manual CyberScan pH 1100/ 2100 54 10.5.8 Memory clear setup mode P4.6 Clears all pH, mV and Ion (pH2100 only) memory points stored in

Page 51 - Figure 169

Instruction Manual CyberScan pH 1100/ 2100 55 11 ELECTRODE CARE 11.1 Electrode Activation DO NOT touch or rub the glass bulb. If you follow the st

Page 52 - MemRecall Setup:P4.3

Instruction Manual CyberScan pH 1100/ 2100 56 12 RS 232 COMMUNICATION This meter will send data to a printer and computer through the RS-232 port

Page 53 - Data Transfer :P4.5

Instruction Manual CyberScan pH 1100/ 2100 57 13 ADDITIONAL INFORMATION 13.1 pH and Temperature The pH electrode is affected by temperature cha

Page 54 - Figure 184

Instruction Manual CyberScan pH 1100/ 2100 58 14 SUMMARY OF DEFAULT SETTINGS / OPTIONS Program Setup Menus Options Factory Defaults pH SETU

Page 55 - 11 ELECTRODE CARE

CyberScan pH1100/ 2100 59 15 TROUBLESHOOTING & ERROR MESSAGES TROUBLESHOOTING Problem Possible Cause Solution Power ON but no display AC

Page 56 - 12 RS 232 COMMUNICATION

Instruction Manual CyberScan pH 1100/ 2100 6 1 INTRODUCTION 1.1 Introducing the Bench Meter Series The pH 1100/ 2100 bench meter is microprocess

Page 57 - 13 ADDITIONAL INFORMATION

CyberScan pH1100/ 2100 60 16 SPECIFICATIONS pH Range -1.000 to 14.000 pH Resolution 0.001 / 0.01 pH (user-selectable) Accuracy ± 0.002 / ±0.0

Page 58

Instruction Manual CyberScan pH 1100/ 2100 61 17 ACCESSORIES Consult your Authorized Distributors for these items and other range of specialized p

Page 59 - Offset of temp. probe

Instruction Manual CyberScan pH 1100/ 2100 62 18 WARRANTY Eutech Instruments/ Oakton Instruments supplies this bench meter with a 3-year warranty

Page 61 - 17 ACCESSORIES

For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or vis

Page 62 - 18 WARRANTY

Instruction Manual CyberScan pH 1100/ 2100 7 1.2 Keypad See Figure 1. A large membrane keypad with touch feedback makes the meter easy to use. Name

Page 63

Instruction Manual CyberScan pH 1100/ 2100 8 1.3 Rear Panel See Figure 2. The pH 1100/ 2100 meters provide a complete set of input connections for

Page 64

Instruction Manual CyberScan pH 1100/ 2100 9 1.4 Electrode Holder The electrode holder is included in the same box as the meter. To attach the e