Transmitter Cond/Res 2-Channel M300 22
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
6.2 Calibration Constant Entry
Enter the calibration constants from the sensor label or certificate. Conductivity cell constants are
primary ‘p’. For 2-electrode sensors, leave A = 0.0000. Temperature constants are secondary
‘s’. When configuring dual channel transmitters pressing [ENTER] will access sensor channel
‘B’ calibration constants. Press [ENTER] again to continue.
6.3 Measurement Units
Select a or b (single channel models) or a or c (dual channel models) and select the units of
measurement. Example:
By selecting a and S/cm as unit, the conductivity value will be displayed on the 1
st
line. By
selecting b and Ω-cm as unit, the resistivity value will be displayed on the 2
nd
line of the display.
By selecting a and °C as unit, the temperature will be displayed on the 1
st
line. By selecting b
and °C as unit, the temperature will be displayed on the 2
nd
line.
6.4 Analog Outputs
On the same screen above, by selecting Yes the linear 4–20 mA analog output Aout1 will be set
up when [Enter] is pressed. Additionally, Aout3 may be configured on dual channel units if
measurement c is selected. Selecting No means that no analog output is set up.
Aout1 min, Aout1 max are the minimum and maximum measurement values for the 4 and
20mA values respectively. Be sure to enter the correct unit multiplier (µ, m, K, M). To configure
additional outputs, use the Configuration Menu. Press [ENTER].