Transmitter Cond/Res 2-Channel M300 45
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
10.3 Mode
(PATH: MENU/PID Setup/Mode)
This menu contains the selection of control modes using relays or analog outputs. Press
[ENTER].
10.3.1 PID Mode
This menu assigns a relay or analog output for PID control action as well as details of their
operation. Based on the control device being used, select one of the following three paragraphs
for use with solenoid valve, pulse input metering pump or analog control.
Pulse Length – If using a solenoid valve, select “Relays” and “PL”, Pulse Length. Choose the
first relay position as #1 if controlling a concentrating reagent feed. Choose the second relay
position as #2 if controlling dilution
water. Set the pulse length (feed cycle time) in seconds. A short pulse length will provide more
uniform feed. A longer pulse length will reduce wear on the solenoid valve. A value of 30 sec-
onds may be a good starting point. The % “on” time in the cycle is proportional to the control
output.
Pulse Frequency – If using a pulse input metering pump, select “Relays” and “PF”, Pulse
Frequency. Choose the first relay position as #3 if controlling a concentrating chemical feed.
Choose the second relay position as #4 if controlling dilution water. Set the pulse frequency to
the maximum frequency allowed for the particular pump being used, typically 60 to100 pulses/
minute. Control action will produce this frequency at 100% output.
CAUTION: Setting the Pulse Frequency too high may cause the pump to overheat.
Analog – If using Analog control, change “Relays” to “Analogout” using up/down arrow keys.
Choose the first Analogout position as #1 if controlling a concentrating chemical feed. Choose
the second Analogout position as #2 if controlling dilution water feed. Select the analog output
current range required by the control device, 4–20 or 0–20 mA. Press [ENTER].
10.4 Tune Parameters
(PATH: MENU/PID Setup/Tune Parameters)
This menu assigns control to a measurement and sets the setpoint, tuning parameters and
non-linear functions of the controller through a series of screens.