Transmitter Cond/Res 2-Channel M300 23
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
6.5 Setpoint
After configuring the Analog Output, a Setpoint can be configured for that measurement. If No is
selected and [ENTER] is pressed then the quick setup is done and the menus are exited without
setting up any Setpoint. To establish a Setpoint for measurement a (and/or c for dual channel
transmitters), select Yes and select one of the following Setpoint Types:
High (High value has to be set)
Low (Low value has to be set)
Between (High and Low value has to be set)
Outside (High and Low value has to be set)
Setpoints for Mettler Toledo Thornton M300 only:
USP (% safety margin below U.S. Pharmacopoeia limits)
EP PW (% safety margin below European Pharmacopeia limits for Purified Water)
EPWFI (% safety margin below European Pharmacopeia limits for Water for Injection)
After setting the Setpoint value(s) and its units multiplier, e.g. µ, m, K, M, select a Relay (none,
1, 2, 3, 4, 5, 6) for that Setpoint. The Relay delay is set to 10 seconds and the Hysteresis is set
to 5%. Press [ENTER].
If [ENTER] is pressed again the Setpoint setup is stored and any previous use of the Relay is
cancelled. If No is selected then the menu is exited without saving the entered settings.