TechnologyMadeEasy...Instruction Manual αlpha pH 190 1/8 DIN pH / ORP Controller with Temperature display and Transmitter 68X276106 R
- 7 - 4 MEASUREMENT MODE The αlpha pH 190 controller is capable of taking pH (or ORP) measurement with Automatic (ATC) or Manual (MTC) Temperature
- 8 - Set Point 2 (SP2) adjustment mode This lets you adjust the pH (or ORP) value in Set Point 2. If this value is crossed, the set point relay 2
- 9 - 5 PASSWORD The Setup mode is to be accessed by entering a password code. The direct accessed Set Point adjustment mode (SP1 and SP2) and th
- 10 - P.P.000P.000P.010P.020P.011P.022P.011P.022
- 11 - 6 PH CALIBRATION IMPORTANT: When Calibration mode is entered, controller automatically goes into a “HOLD” mode where the 4-20 mA output fre
- 12 - NOTE: If the ATC mode has been selected and the Pt 100 is not connected or if it is broken, the pH display will flash. Under this condition, th
- 13 - If you are performing a 1-point calibration, press the ▲ and ▼keys together to return to measurement mode. If you are performing a two-point
- 14 - 7 ORP CALIBRATION IMPORTANT: When Calibration mode is entered, controller automatically goes into a “HOLD” mode where the 4-20 mA output freez
- 15 - 8 TEMPERATURE CALIBRATION IMPORTANT: When Calibration mode is entered, controller automatically goes into a “HOLD” mode where the 4-20 mA o
- 16 - 9 SETUP MODE 9.1 General Information IMPORTANT: When Setup mode is entered, the controller automatically goes into a “HOLD” mode where the 4-
Preface This manual serves to explain the use of the αlpha pH 190 controller/transmitter. This manual functions in two ways: first, as a step-by-st
- 17 - 9.2 Setup mode overview P1.0: Set Point 1 P1.1: select relay 1 set point value P1.2: select relay 1 as low or high set point P1.3: set
- 18 - 9.3 Set Point 1 – P1.0 Setup program P1.0 allows you to set parameters for relay 1. P1.1: select relay 1 set point value P1.2: select relay 1
- 19 - P1.3: Set Hysteresis value Hysteresis prevents rapid contact switching if your value is fluctuating near the set point. Once activated, the
- 20 - 9.4 Set Point 2 – P2.0 Setup program P2.0 allows you to set parameters for relay 2. P2.1: select relay 2 set point value P2.2: select relay 2
- 21 - P2.3: Set Hysteresis value Hysteresis prevents rapid contact switching if your value is fluctuating near the set point. Once activated, the
- 22 - 9.5 Configuration – P3.0 Setup program P3.0 let you configure controller parameters. P3.1: select pH or ORP mode of operation P3.2: select sym
- 23 - P3.3: Select temperature compensation This controller features selectable automatic temperature compensation (ATC) or manual temperature co
- 24 - 9.6 Viewing Calibration points: pH or ORP – P4.0 Setup program P4.0 lets you view the current points at which the controller is calibrated. Th
- 25 - 9.7 Viewing pH Electrode Data – P5.0 In pH mode, Program 5 has three “view only” options that lets you check the electrode parameters for d
- 26 - 9.8 Viewing ORP Electrode Data – P5.0 In ORP mode, Program 5 has two “view only” options that lets you check the electrode parameters for diag
TABLE OF CONTENTS 1 INTRODUCTION ...1 2 SAFETY INFORMATION
- 27 - 9.9 Controller Reset – P6.0 This parameter lets you reset the controller. All pH and ORP calibration data will be cleared, and all Setup pa
- 28 - 10 RELAYS The αlpha pH 190 features two SPDT non-powered relays; rated for 6A at 110 VAC, 250 VAC maximum. When your process exceeds the set pa
- 29 - 13 SPECIFICATIONS pH Range 0.00 to 14.00 pH Resolution/Accuracy 0.01 / ± 0.01 pH mV Range -999 to 1000 mV Resolution/Accuracy 1 / ± 2 mV
- 30 - 14 ACCESSORIES Product Description Code no. pH Combination Electrode with Pt 100 RTD and PMP EC-100GTSO-05B pH Combination Electrode with HF r
- 31 - Appendix 1: Asymmetrical / Symmetrical Mode The αlpha pH 190 controller can be operated in the asymmetrical or symmetrical modes (See Set u
- 32 - Symmetrical mode: For noisy electrical environments, the alpha pH 190 controller offers Symmetrical Mode operation. To take advantage of Symm
- 33 - Appendix 2: Simple Explanation on the Function of Hysteresis The controller relay activates when the set-point is reached. In the reverse
- 34 - Appendix 4: External Relays The relays on the αlpha pH 190 series controller are rated for 6 amps at 110 VAC and can be wired directly to your
- 35 - 15 GENERAL INFORMATION Warranty Eutech Instruments warrants this product to be free from significant deviations in material and workmanship
For more information on Eutech Instruments products, contact your nearest Eutech Instruments distributor or visit our website list
- 1 - 1 INTRODUCTION Thank you for purchasing the αlpha pH 190 ⅛ DIN pH/ORP Controller. This controller is part of a series of quality process con
- 2 - 2 SAFETY INFORMATION The Eutech Controller/Transmitter shall be installed and operated only in the manner specified in the Instruction manual.
- 3 - 3 OVERVIEW 3.1 Front Panel The front panel consists of a 4-digit LED display, 8 LED annunciators and 4 keys. Annunciators 1. REL 1 Displa
- 4 - 3.2 Back Panel The back panel consists of four different connectors that can be used with removable terminal blocks (included): + 4 - 20 mA12
- 5 - 3.3 Wiring Caution: Ensure electrical mains are disconnected before proceeding. 1. Connect the power supply to the three-pin terminal blo
- 6 - 3.4 Panel-mounting the controller The supplied mounting hardware allows surface mounting to all panels and protective enclosures. Mounting cut-