EUTECH INSTRUMENTS ALPHA RES 1000 RESISTIVITY CONTROLLERTRANSMITTER Instruction Manual Page 1

Browse online or download Instruction Manual for Noise Reduction Machine EUTECH INSTRUMENTS ALPHA RES 1000 RESISTIVITY CONTROLLERTRANSMITTER. EUTECH INSTRUMENTS ALPHA RES 1000 RESISTIVITY CONTROLLERTRANSMITTER Instruction manual User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Alpha pH 2000

Instruction Manual Alpha pH 2000 pH/ORP Controller Transmitter pH / mV / ºC meter 68X216830 | Rev 3 | Apr 2008

Page 2

3 2 PRODUCT DESCRIPTION 2.1 Description of unit The Alpha pH 2000 is used for measuring pH or ORP and temperature values. The pH or ORP values can b

Page 3

4 2.2 Measurement and Control System A typical measurement system consists of: • a pH/ORP process controller • a pH/ORP combination electrode wi

Page 4

5 2.3 Unit Overview Wall mounting version Panel mounting version

Page 5 - TABLE OF CONTENTS

6 2.3.1 Display overview The LC display shows two alpha-numerical fields for parameters and measured values as well as various mode and status indica

Page 6

7 LED indicators Relay indicators If REL key is pressed the LED (A, B or W) indicates to which Relay (A, B or Wash) the displayed limit values r

Page 7

8 SETUPHOLDSETUPHOLDSETUPHOLDSETUPHOLDSETUPHOLDSETUPHOLDcur.1SETUPHOL Dcur.2SETUPHOLDSETUPHOLDMEASPHOCSETUPHOLDTemperature settingssee section 6.3Rel

Page 8 - 1.2 Intended use

9 3 ASSEMBLY AND INSTALLATION 3.1 Mounting the unit Wall Mounting Version approx. 14 [.55]39 [1.54]Holes for wallmounting (2X)21.5 [.85]6 [.24]

Page 9 - 1.3 Safety instructions

10 Panel mounting version Transmitter housing for panel mounting: protection class IP 54 (front), IP 40 (housing) Flat gasket 1mm [.04](to be inse

Page 10 - 2.1 Description of unit

11 3.2 Connection Diagram Caution: Ensure electrical mains are disconnected before proceeding. Connection for wall mounting version RELA

Page 11 - Flow Assembly

12 NOTE: Switch or circuit breaker shall included in the building installation. It shall be in close proximity to the equipment and within easy

Page 13 - 2.3.2 Key functions

13 Connection for panel mounting version 1. AC mains live wire 17. no connection 2. AC mains neutral wire 18. no connection 3. AC mains pro

Page 14

14 4 NORMAL OPERATION 4.1 Measurement mode When the controller is powered on, the display first shows all segments briefly, then the transmitter au

Page 15 - Menu overview

15 5 CALIBRATION MODE You can access the Calibration mode directly from the Measurement mode by pressing the CAL key and entering the Calibration sec

Page 16 - 3 ASSEMBLY AND INSTALLATION

16 5.2 pH Calibration This controller unit features one-point or two-point calibration with five preset buffer sets. Buffer values refer to 25 °C. T

Page 17 - UNIT: MM [INCH]

17 7. Press the ENT key to return to the Measurement mode. Note: If you entered the Calibration mode from the Setup mode, the controller will return

Page 18 - Connection Diagram

18 5.4 ORP -% Calibration If controller is configured for ORP-% measuring you can perform a calibration at two points using a low concentration samp

Page 19

19 5.5 View actual zero and slope ENTENTCALPHCALMEASPHOCatc 1. While in Measurement mode press the CAL key. The display will prompt you to enter a

Page 20

20 MEASPHOCENTERENTENTERENTENTERENTENTERENTENTERENTENTERENTENTERENTENTERENTENTERENTENTERENTENTERENTENTERENTSETU PHOLDSETUPHOLDSETUPHOLDSETUPHOLDSETUP

Page 21 - 4.1 Measurement mode

21 6.2 Electrode Offset (OFS) sub-function The electrode offset sub-function is only available if the controller is configured for pH measuring. Use

Page 22 - 5 CALIBRATION MODE

22 6.3 Setting temperature (Set oC) sub-function SETU PHOLDSETUPHOLDOCSETUPHOLDOSETUPHOLDSETUPHOLDOCSETUPHOLDOCatcSETUPHOLDSETUPHOLDENTENTENTENTENTE

Page 23 - 5.2 pH Calibration

Preface This manual serves to explain the use of the Alpha pH 2000 series pH/ORP controller/transmitter. The manual functions in two ways, firstly as

Page 24 - 5.3 ORP–mV Calibration

23 6.4 Control Relay A/Control Relay B (SP1/SP2) sub-function The SP1 sub-function determines the operating parameters for Relay A; while SP2 define

Page 25 - 5.4 ORP -% Calibration

24 the relay for the number of seconds (0 to 1999 seconds) you select. Press the ENT key to confirm your setting. Note: You can set a time delay

Page 26 - View actual zero and slope

25 6.5 Controller (CNTR) sub-function The CNTR sub-function determines the controller’s parameters. 7 8991011125SETUPHOLDSETUPHOLDSETUPHOLDSETUPH

Page 27 - Enter Setup mode

26 - OFF = controller off Use control Off to operate controller as a monitor only or to prevent relays from switching. - L.CT = limit value control

Page 28

27 10. Setting the proportional range: press the ▲ or ▼ key to set the proportional range (setting range: 10 to 500%). Press the ENT key to confirm

Page 29 - C) sub-function

28 Current Output 2 sub-function In this sub-function you set the current output range of the transmitter for temperature values. ENTENTENTENTENTEN

Page 30

29 6.7 Wash relay (WASH) sub-function In this sub-function you set the parameters for the wash relay. 1223345ENTENT ENTENTENTENTENTSETUPHOLDSETUPH

Page 31

30 6.8 Configuration (CNFG) sub-function In this sub-function you configure the controller to suit your requirements. SETUPHOLDSETUPHOLDSETUPHOLDS

Page 32

31 – “PH SY” = pH measurement with symmetrical input – “PH ASY” = pH measurement with asymmetrical input – “ORP % SY” = ORP-% measurement with symm

Page 33

32 8. Resetting the controller settings to factory defaults: The display shows “NO DEF”. Press the ▲ or ▼ key to select: – “NO DEF” = keeps old value

Page 34

Safety Information The Alpha pH 2000 Controller/ Transmitter shall be installed and operated only in the manner specified in the Instruction manual.

Page 35

33 7 RELAY MODES You can control devices connected to Relay A, Relay B or wash relay via the front panel of the transmitter. In Automatic mode, the c

Page 36

34 7.2 Manual relay mode In manual relay mode, you can manually turn “on” and “off” the control devices connected to Relay A, Relay B or Wash relay.

Page 37

35 8 TECHNICAL SPECIFICATIONS 8.1 General Specification Alpha pH 2000 Controller / Transmitter pH Range -2.00 to 16.00 pH Resolution & Accuracy

Page 38

36 EMC Specifications Emitted Interference EN 61 326 Immunity to Interference EN 61 326 Environmental Conditions Ambient Temp. Operating Range -10

Page 39

37 8.3 Specifications for panel mount version Electrical Data And Connections Signal output Two 0/4 to 20 mA outputs for pH/mv and temperature , ga

Page 40 - 7.1 View relay set points

38 9 ACCESSORIES 9.1 Replacement Unit Product Description Thermo Scientific Order Code Alpha pH 2000 pH/ORP Controller/Transmitter, wall mount ve

Page 41 - 7.2 Manual relay mode

39 10 GENERAL INFORMATION 10.1 Warranty This Controller transmitter is supplied with a one-year warranty against significant deviations in materia

Page 42 - 8.1 General Specification

40 10.4 Guidelines for Returning Unit for Repair Use the original packaging material if possible when shipping the unit for repair. Otherwise wrap i

Page 43

41 11 APPENDICES 11.1 Appendix 1 – Unit fuse and jumper settings − Caution! Before opening the unit to replace the unit fuse or to set the jumpe

Page 44

42 Panel mounting version J4 J4R44C55C59REG5REG1C69C66C68R49R48R46D10C73R45REG7C107C124C125R107C123U8R94C104C91C92C85C89C88R77C90Q11Q10C86C84R79C87

Page 45 - Assembly Accessories

TABLE OF CONTENTS 1 INTRODUCTION ...

Page 46 - Return of Goods

43 11.2 Appendix 2 – pH buffer values at various temperatures The following table shows the various pH values at different temperature of the solut

Page 47 - Maintenance and Cleaning

44 11.3 Appendix 3 –Simple explanation on the function of hysterisis 4 . 0 4.5 9.5 10.0 7.0R E L A Y O F F H Y S T E R E S I S B A N D ( D E F A U

Page 48 - 11 APPENDICES

45 11.4 Appendix 4 – General instructions concerning Controller Setting 11.4.1 Control characteristic of Controllers used as limit value switch MI

Page 49 - Panel mounting version

46 11.4.4 Controller signal of Pulse length Controllers tONtOFFPulse Length TRelayONOFFTime [s] The output relay of the pulse length controller is c

Page 50

47 11.4.5 Controller signal of Pulse Frequency Controllers tONtOFFPulse Length TRelayONOFFTime [s] The output relay of the pulse frequency controlle

Page 51 - 4 . 0 4.5 9.5 10.0

48 11.5 Appendix 5 – Abbreviations used in menu displays Abbreviation Meaning Abbreviation Meaning MEAS Measurement CUR.1 Output current 1 CAL

Page 52

For more information on our products, contact your nearest distributor or visit our website listed below: Thermo Scientific W

Page 53 - Time [s]

10.3 RETURN OF GOODS ... 39 10.4 GUI

Page 55 - 4 - 20 4 to 20mA

1 1 INTRODUCTION 1.1 At the very beginning We thank you for having purchased the Alpha pH 2000 pH/ORP Controller. The construction of the pH/ORP Co

Page 56

2 1.3 Safety instructions − The Alpha pH 2000 should be installed and operated only by personnel familiar with the transmitter and who are qualifi