3. DISSOLVED OXYGEN / ATC ELECTRODE The DO 700 includes a galvanic DO electrode with built in temperature measurement. The electrode has a dual cab
5 4. CALIBRATION The DO 700 has three measurement modes; DO as % saturation, DO as mg/L (or ppm) concentration, and temperature. Dissolved oxygen l
The DO 700 will accept one calibration point in concentration mode. The minimum value is 2 mg/L (ppm), and the calibration window is +/- 40% of the fa
4.2 % Saturation Calibration The DO700 can be easily calibrated in air. For best accuracy, ensure that the barometric pressure value is accurate. Th
4) Press . The CAL indicator will blink and the secondary display will indicate the calibration point, 0.0. 5) Press to confirm the calibration. Th
9 NOTE: The minimum calibration value is 2 mg/L. In addition, the calibration is limited to ±40% adjustment of the factory default value to prevent
5. SETUP FUNCTIONS Use the setup feature to customize your instrument operation. There are two setup menus; one menu is derived from the % saturation
Press to view Offset % (OFS)—% saturation mode only. 5.3 3.0 ConF (Configuration) Press to access set-up programs 3.1 thru 3.5. 5.4 3.1 rdY (Re
5.7 3.4 SALt (Salinity Adjustment) For best accuracy, enter the actual salinity value if your samples are at least 0.1 ppt (100 ppm). Maximum adjust
5.10 4.0 rSt (Reset) Press . Press or to select “Yes” (Reset) or “No” (Cancel). If “Yes”, press or to select “Cal” (calibration reset only) or “
6. TAKING MEASUREMENTS During measurement the dissolved oxygen reading is automatically compensated for pressure and salinity. For best accuracy, ens
6.1 HOLD Function For prolonged observation of a reading, press during measurement mode to freeze the display. The “HOLD” indicator will display whe
16 7. ELECTRODE MAINTENANCE The DO 700 electrode produces an output proportional to the oxygen present in the medium in which it is placed. The gal
There are no special probe storage requirements. Cap and Electrolyte Replacement Replacement of the pre-assembled cap is required only when you canno
18 Electrolyte Solution The electrolyte solution in your probe’s cap will deplete on usage and will need to be replaced periodically. The replacemen
8. DISSOLVED OXYGEN THEORY Dissolved Oxygen (DO) refers to the volume of oxygen that is contained in water. There are two main sources of DO in wate
20 Measurement Units One measure of DO in water is parts per million (ppm) which is the number of oxygen molecules (O2) per million total molecules
21 Air Calibration Understanding the principle of air calibration is easy, once you know that it is partial pressure that the probe is responding t
22 Some pollutants such as acid mine drainage produce direct chemical demands on oxygen in the water. DO is consumed in the oxidation-reduction rea
23 9. REPLACEMENTS AND ACCESSORIES Part number Ordering Code Item Description Eutech Instruments Oakton Instruments DO 700 meter with DO/ATC probe
10. TROUBLESHOOTING GUIDE PROBLEM CAUSE SOLUTION No display Main power not switched on. AC Adapter socket not inserted properly. Switch on the powe
PROBLEM CAUSE SOLUTION CAL % calibration value is not acceptable. Electrode is disconnected or failing. Check electrode connections. Check that c
26 11. SPECIFICATIONS DO % Saturation Range Resolution Relative accuracy Offset adjustment Calibration Points 0 to 300% 0.1 % (0.0 to 199.9%);
27 12. WARRANTY This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from
28 13. RETURN OF ITEMS Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any
For more information on our products, please contact our channel partner or visit our websites listed below: Oakton Instruments 625 E Bunker
TABLE OF CONTENTS 1. INTRODUCTION ... 1 2. GETTING STARTED...
9. REPLACEMENTS AND ACCESSORIES... 23 10. TROUBLESHOOTING GUIDE... 24 11. SPECIFICATION
1 1. INTRODUCTION Thank you for purchasing our DO 700 series benchtop meter. This microprocessor-based meter is economical and simple to use. The d
2. GETTING STARTED 2.1 Keypad Functions Powers the meter on and off. Upon power on, the meter automatically begins in the mode that was last used.
2.2 LCD Annunciators 2.3 Meter Connections DC Power supply ATC Temperature connection of Dissolved Oxygen electrode DO BNC connection of Disso