Display of the VibraCheck II:
1. Bias Indicator: The Bias indicator is located on the left side of the display. The VibraCheck Limited and
VibraCheck II continuously monitors the bias condition of the ICP accelerometer and indicates the status on the
display. There are 3 conditions: OK, High and Low.
OK: ICP bias is in the normal range, everything is working correctly including the cable.
High: Usually the sensor or cabling is not connected to the meter.
Low: The ICP circuit is shorted. Check the cabling immediately.
Caution: Do not operate the meter when you observe a high or low bias condition. Inspect and replace as
required.
1. Bias Indicator status:
OK-
Open-
Short-
4. Battery Indicator
3. Vibration units and
method of detection
2. Numerical value of the
vibration level
5. Hold, Gain,
Average or Peak and
memory review modes –
V=Pod II only
6. Storage location label for point storage or
route based data collection storage- V-Pod II
only
7. Stored data indicator
V-Pod II only
Low
High
OK