Instruction manual αlpha DO 2000
26
6.7 Current Output 2 sub-function
In this sub-function you set the current output range of the transmitter for temperature values.
1. Select the “CUR.2” sub-function, then press the ENT key.
2. Selecting the output type: press the or key to select the desired output type: 0-20 mA
or 4-20 mA. Press the ENT key to confirm your selection.
3. Setting the temperature value at which the transmitter output will be 4 mA (or 0 mA):
press the or key to set the temperature value (setting range: -10.0 to 115.0 °C or 14 to
237 °F) to be equivalent to 4 mA (or 0 mA). Press the ENT key to confirm your setting.
4. Setting the temperature value at which the transmitter output will be 20 mA: press the
or key to set the temperature value (setting range: 0.0 to 125.0 °C or 32 to 257 °F) to be
equivalent to 20 mA. Press the ENT key to confirm your setting.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
and keys simultaneously (escape).