6.3 Normalization Temperature
You can set the meter to normalize its conductivity measurements to a standard
temperature of either 25 °C or 20 °C.
The default value is 25 °C.
1) Enter the advanced setup as described in
Section 6.1.
2) Press or key until [t.nr °C] appears
on the LCD. Press ENTER key.
3) Press or key to select either [25.0
°C] or [20.0 °C].
4) Press ENTER key to select. The meter
will take you back to the menu, [t.nr °C].
5) Press or key to move to the next
menu or press CAL to exit to
measurement mode.