EUTECH INSTRUMENTS ALPHA CON 500 2-WIRE TRANSMITTER Instruction Manual Page 35

  • Download
  • Add to my manuals
  • Print
  • Page
    / 56
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 34
Instruction Manual αlpha CON 2000
29
6.8 Wash relay (WASH) sub-function
In this sub-function you set the parameters for the wash relay.
SETUP
SETUP
HOLD
HOLD
4
4
ENT
SETUP
SETUP SETUP
HOLD HOLD HOLD
444
ENT
ENT ENT
ENT
ENT
SETUP
SETUP
HOLD
HOLD
4
4
1. Select the “WASH” sub-function, then press the ENT key.
2. Enabling/disabling wash function: press the or key to enable (WASH ON) or
disable (WASH OFF) wash function. Press the ENT key to confirm your selection.
3. Selecting the relay status condition: press the or key to choose desired relay
status (de-energised = “DEEN” or energised = “EN”). Press the ENT key to confirm your
selection.
4. Setting the wash interval in hours: Press the or key to set the desired wash
interval (setting range: 0.1 to 199.9 hours). Press the ENT key to confirm your setting.
5. Setting the wash duration in seconds: Press the or key to set the desired wash
duration (setting range: 1 to 1999 seconds). Press the ENT key to confirm your setting.
Note: During wash cycle the transmitter is set to HOLD. For safety reasons, the
HOLD function is activated 5 seconds prior and deactivated 10 seconds after the
wash cycle.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
and keys simultaneously (escape).